HT2800T_en_c.doc Page 260 of 275
•
Set the Bar Code Reader Module HT2800T in order to activate only the reading of the
Code type used (for example if Code 128 is being used, enable only this code type in
order to reduce the possibility of reading error).
•
Follow the indication of the Bar Code standards when you create it. In particular leave an
appropriate margin on the left and on the right of the Bar Code.
12.1.2.4 Cap specifications
The autosampler uses sample vials with crimp caps.
The Minimum diameter of the pierceable area is 9mm.
In general, do not use crimp caps more than once for Headspace analysis.
12.1.2.5 Septa specifications
Septum material
Compatible with
Incompatible with
Thickness
PTFE/silicone
rubber
PTFE resistance until
punctured, then septa
will have the
compatibility of silicone
(alcohol, acetone, ether,
DMF, DMSO)
ACN, THF, benzene,
chloroform, pyridine,
toluene, hexane,
heptane
2-3mm
For more detailed compatibility data refer to the manufacturer's recommendations.
In general, do not use septa more than once for Headspace analysis.
12.2 Consumables for Liquid mode
12.2.1 Syringes
Note: The autosampler supports a wide range of syringes. Ensure the syringe is compatible with
your application and analyzer. If in doubt, contact your Technical Representative.
HTA
part
number
Syring
e size
Needle
kind
Plunger
with PTFE
cover
Needle
length
(mm)
Syringe
holder
Plunger
block
position (*)
Gauge
Notes
1.23.105
10ul
Fixed,
conical tip
No 80 1 S G22
1.23.106
10ul
Fixed,
conical tip
No 80 1 S G23