HT2800T_en_c.doc Page 261 of 275
To use the syringeID system (included in the GLP pack, see 1 . 7 . 2 Options”) you have
to install the following syringe:
HTA part
number
Syringe
size
Needle
kind
Plunger
with PTFE
cover
Needle
length
(mm)
Syringe
holder
Plunger
block
position (*)
Notes
1.91.759
10ul
Removable,
conical tip
No 80 Included
in the
GLP pack
S
With
SyringeID
(*) “B” and “S” are the two sides of the plunger block (see chapter 2.1.2)
Spare parts for 1.91.759:
Needle (80mm): 1.23.116.
Warning
The syringe information are loaded in a RFID tag included in the black adapter mounted on
the syringe for the syringeID system (see paragraph 11.2 “Liquid mode“). Every adapter must
be associated to a syringe only in order to maintain the info regarding the syringe always
correct and updated. Do not remove the syringe from its black adapter.

Table of Contents