HT2800T_en_c.doc Page 214 of 275
8.9 Mismatch errors
8.9.1 Syringe wash mismatch error
8.9.1.1 Syringe mismatch
This error appears only in Liquid mode, if SyringeID system is enabled (see paragraphs 11
“Appendix A – Glossary“ and 4.2.6.8 “Setup configuration: SyringeID“).
Error message:
“Syringe mismatch! Execute Syringe Installation”
Causes:
At the beginning of a syringe wash, the syringeID system detects the presence of a syringe
different from the one detected during the syringe installation procedure (see paragraph
4.2.3Syringe installation/ replacement (liquid mode).
Solution:
Tap “ABORT” and perform the syringe installation procedure (see paragraph 4.2.3 “Syringe
installation/ replacement (liquid mode). Then restart the washing operation.
8.9.2 Single injection mismatch error
If the method parameters conflict with those defined in “setup”, the unit will display an error
message.
Examples are shown below.
8.9.2.1 Syringe Volume mismatch
Error message:
Warning: Run Error Parameters” “Method X: Syringe Volume mismatch”
(Where “X” is the method used for the single injection)
Causes:
The syringe volume set during installation is different from the syringe volume set in the method.
Solution: