8.2.3 Memory error
8.2.3.1 BMO reading error
Error message:
“Error during BMO reading!"
Causes:
Autosampler cannot read the Border Mark Out parameters.
Solution:
Tap “CONTINUE”. The “Setup: Service” screen appears (see paragraph 4.2.7”Setup: Service“)
Contact your Supplier or the Technical Assistance Centre to solve the problem.
8.2.3.2 Method reading error
Error message:
“Error during methods reading!"
Causes:
Autosampler cannot read the parameters of the saved methods.
Solution:
Tap “CONTINUE”. Try to switch on-off the autosampler. If the error message appears again,
contact your Supplier or the Technical Assistance Centre to solve the problem.
8.2.3.3 Sequence reading error
Error message:
“Error during sequences reading!"
Causes:
Autosampler cannot read the parameters of the saved sequences.
Solution:
Tap “CONTINUE”. Try to switch on-off the autosampler. If the error message appears again,
contact your Supplier or the Technical Assistance Centre to solve the problem.