HT2800T_en_c.doc Page 60 of 275
3.3.5.3 Cleaning
Remove the vials (see above).
Carefully remove the plastic caps and septa.
Empty and wash the glass vials and the plastic caps.
If necessary, change the septa.
Refit the caps and septa on the vials as described in the “Mounting“ (see above).
Replace in their original place (see
Figure
1 and Figure 5);
3.3.5.4 Filling the Solvent Vials
Remove the vials (see above).
Carefully remove the plastic caps and septa.
Empty and wash the glass vials and the plastic caps.
If necessary, change the septa.
Fill the vials with the solvent.
Refit the caps and septa on the vials as described in the “Mounting“ (see above).
Put back the vials in their original place (see
Figure
1 and Figure 5).
3.4 Start-Up
The main power switch is on the back of the autosampler (see Figure 30, switch 1), switch it ON (“I”
position).
When the autosampler is switched on a program loading screen appears (see paragraph 2.4.2Limited
Warranty). Then the displays shows the autosampler firmware release.
After the program has loaded the display shows the “HOME SCREEN” (see paragraph 7.1 Home
screen“).
During the start up process, in Headspace and SPME mode, the autosampler will proceed to check
the incubation oven. If the oven is not empty, a screen appears asking if the oven has to be unloaded. If
so, a following screen asks where to move the vial(s).