“Missing: touch & plunger zero”/ “Missing: Fiber zero”.
Causes:
“Touch Plunger Zero” procedure is missing (4.2.1.4 Alignment: Touch Plunger Zero (Headspace
and Liquid mode)/ Touch Fiber Zero (SPME mode)”.
Solution:
Execute the “Touch Plunger Zero” procedure/ Fiber zero.
8.11 Timeout error
8.11.1 Peripheral communication timeout
Error message:
“Time-out communication with peripheral: X”
(Where “X” can be Plunger, Needle, Turret, Tray, Shaker etc...).
Causes:
The communication with the a peripheral fails.
Solution:
Tap “RETRY” to retry the communication with the peripheral.
Tap “DISENGAGE” to enter in the screen from which it is possible to move the motors to try to
unlock the system and to solve the problem (4.2.5 “Setup: Manual operations).
Tap “ABORT to stop the operations.
If the problem isn't solved, switch the autosampler OFF, wait 30 seconds and switch it ON again.
Retry the operation.
If the problem persists contact the Technical Assistance Centre, supplying the error message.
8.12 SyringeID timeout
This error appears only in Liquid mode, if SyringeID system is enabled (see paragraphs
1 1 Appendix A – Glossary” 11 a n d 4 . 2 . 6 . 8 Setup configuration: SyringeID “).
Error message: SyringeID timeout!”
Causes:
The syringeID parameter reading by the SyringeID system is not successful during the
SyringeID parameter reading test (see paragraph 4 . 2 . 6 . 8 Setup configuration: SyringeID” ) or
at the beginning of a syringe wash, a single injection or an automatic run.
Solution:
Tap “RETRY” to retry the syringeID parameter reading. Tap “ABORT” to abort the running
operation.