HT2800T_en_c.doc Page 175 of 275
Warning
If syringeID system is used, be careful while handling the syringe holder of the SyringeID
system (1.7.2 “Options” and 11.2 “Liquid mode”). Do not touch the connector or the
electronic circuit placed on the back of the holder. The components may be damaged. If
not used, the SyringeID holder must be stored in the antistatic bag supplied with the
packaging.
7.10.1 Change from Headspace to Liquid mode
Follow this procedure to move from Headspace to Liquid mode
Access sequence: RUN>CHANGE MODE>LIQUID
The screen “You are moving to Liquid mode” confirms the start of the changing procedure. Tap
“CONTINUE"” to start the switching operation or “ABORT” to return to the “CHANGE MODE”
screen.
Warning
If syringeID system is used, the syringe information are loaded in a RFID tag included in
the black adapter mounted on the syringe for the syringeID system (see paragraph 11.2
“Liquid mode”). Every adapter must be associated to a syringe only in order to maintain
the info regarding the syringe always correct and updated. Do not remove the syringe
from its black adapter.