HT2800T_en_c.doc Page 217 of 275
Restart the automatic run.
8.9.3.4 Syringe mismatch
This error appears only in Liquid mode, if SyringeID system is enabled (see paragraphs 11
“Appendix A – Glossary“ and 4.2.6.8 “Setup configuration: SyringeID“).
Error message:
“Syringe mismatch! Execute Syringe Installation”
Causes:
At the beginning of the automatic run, the syringeID system detects the presence of a syringe
different from the one detected during the syringe installation procedure (see paragraph
paragraph 4.2.3 “Syringe installation/ replacement (liquid mode)”.
Solution:
Tap “ABORT” and perform the syringe installation procedure (see paragraph paragraph 4.2.3
“Syringe installation/ replacement (liquid mode)”. Then restart the automatic run.
8.10 Missing errors
8.11.1
Missing: front injector alignment
Error message:
“Missing: front injector alignment”
Causes:
Front injector alignment (4.2.1.1 “Alignment: Analyzer Default”) has not been performed after the
change of Analyzer brand or model (4.2.1.1 “Alignment: Analyzer Default “)
Solution:
Perform the Front injector alignment (4.2.1.2 “Alignment: Front Injector“).
8.10.1 Missing: rear injector alignment
Error message:
“Missing: rear injector alignment”
Causes:
Rear injector alignment 4.2.1.3 “Alignment: Rear Injector” has not been performed after the
change of Analyzer brand or model (4.2.1.1 “Alignment: Analyzer Default “).
Solution:
Perform the Rear injector alignment (4.2.1.3 “Alignment: Rear Injector”).
8.11.3 Missing: touch & plunger zero/Fiber zero
Error message: