HT2800T_en_c.doc Page 216 of 275
change the syringe volume set during installation if this doesn't match the installed
syringe.
Then restart the automatic run.
8.9.3.2 Tray Type mismatch
Error message:
Warning: Run Error parameters” “Step X: Tray Type mismatch” (Where “X” is the sequence step)
Causes:
The tray type set in the TRAY TYPE” screen (see paragraph 7.9.1 Change tray/rack type”) is
different from that set in the sequence step (see paragraph 6.1.2 Setting sequence step
parameters“).
Solution:
Tap “EXIT” and/or:
change the tray type set in the sequence step to match the installed tray (see paragraph 6
“Programming Sequence“);
change the tray type set in the “TRAY TYPE” screen (see paragraph 7.9.1 Change tray/rack
type “) to match the installed tray.
replace installed tray with the correct one (see paragraph 4.2.1.5 Alignment: Tray Install”) and
change the tray type set in the “TRAY TYPE” screen (see paragraph 7.9.1 Change tray/rack
type”.
Then restart the automatic run.
8.9.3.3 Injection Mode mismatch
Error message:
Warning: Run Error parameters” “Step X: Injection Mode mismatch”
Causes:
The injection mode in the sequence step (see 6.1.2 Setting sequence step parameters“) is
different from the injector type/number in the setup (see paragraph 4.2.1 Setup: Alignment”).
Example: only one injector is defined in setup, but “Rear” or “Confirmation” injection mode is
used in the sequence step.
Solution:
Tap “EXIT”. Modify the injection mode in the sequence step (see paragraph 6.1.2 Setting
sequence step parameters“) or the injector type/number in setup (see paragraph 4.2.1 Setup:
Alignment”). In the latter case run the injector alignment.