HT2800T_en_c.doc Page 212 of 275
To reduce this problem, the crimping tool should be regularly checked to ensure that the cap
cannot turn once it is crimped, and that the under edge of the cap is flush with the vial lip. Also,
check that any obstacles do not hinder the turret and that the vial handlers at the end of the arm
are clean.
In that case we recommend to check:
check the vial handlers are not broken;
check the vial, septa and the crimp cap;
check if the turret has had an unexplained collision with some object, and if vial handlers
or the vials are dirty.
check the turret alignment, if not correct, contact your Supplier or the Technical
Assistance Centre.
8.6 Obstacle found
Error message:
Warning: Obstacle found over the X!“Try to remove the obstacle”
(Where “X” can be a sample vial, a position in the oven or the front or rear Injector)
Causes:
The autosampler has found an obstacle over the indicated position.
Solution:
Remove the obstacle. If necessary, tap “DISENGAGE” to enter the manual operation screen to
move the motors and unlock the system (see paragraph 4.2.5 “Setup: Manual operations“).To
exit manual operations, tap “EXIT”.
Tap “RETRY” to re-start the Automatic run procedure from the point of interruption.
If the problem persists, the Touch & Plunger procedure must be repeated 4.2 . 1.4
Alignment: Touch Plunger Zero (Headspace and Liquid mode)/ Touch Fiber Zero (SPME mode)”.
Warning
The autosampler only works properly by using vials with a round bottom. It has been tested
with vial types reported in 12.1.2 “Sample vials “.