HT2800T_en_c.doc Page 123 of 275
•
Touch sensor: it is “ON” if the vial locator has detected the presence on an object, or
“OFF” if it is not touching.
•
Shaker sensor: it is “ON” when the shaker is in the position suitable for vial loading and
unloading, it is “OFF” in all the other positions.
Ready: shows the “Ready” signal status (see paragraph 0 “Select “Exit” to exit from
this screen and return to the “Setup Configuration”” screen (see paragraph 4.2.6 “Setup:
Configuration”):
•
Setup configuration: Input Output “).
Start: shows the “Start” signal status (see paragraph 0 “ Select “Exit” to exit from this
screen and return to the “Setup Configuration”” screen (see paragraph 4.2.6 “Setup:
Configuration”):
•
Setup configuration: Input Output “). By pressing “OK”, this signal is activated for about 1
second.
•
Sync-out: shows the “Sync-out” signal status. Pressing “OK” this signal is activated.
•
Sync-in: show the “Sync-in” signal status.
•
Flush: shows the status of the flushing valve. It is “ON” when the valve is open and the
gas can flush the syringe, it is “OFF” when the valve is closed. Press “OK” to change the
valve status.
Select “Exit” to exit from the “Setup Manual Operation” screen and return to the “Setup” screen
(see paragraph 4 “Set Up operations “).
4.2.6 Setup: Configuration
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION
This is used to define some general settings of the autosampler. The information is shown on
multiple pages.
Use this screen to access the following sub menus:
•
Sound: changes the sound effects of the autosampler (see paragraph 4.2.6.1
“Setup configuration: Sound”); “
•
Light: changes the target illumination settings (see paragraph 4.2.6.2
“Setup configuration: Light“);
•
Run: changes the settings for the injection sequence (see paragraph 4.2.6.3
“Setup configuration: Run“);
•
Input Output: changes the input/output parameters to/from the analyzer (see paragraph