HT2800T_en_c.doc Page 233 of 275
Dry with a dry cloth.
Replace the rack inside the tray and close the tray.
9.1.2 External cleaning of cabinet
Switch OFF the autosampler.
Remove the power cable from the rear.
Carefully clean the control panel (touch screen or keypad) and fragile parts using
only a damp cloth. They must not be cleaned with solvents or detergents.
Clean the cabinet using a non-abrasive detergent solution.
Dry with a dry cloth.
Do NOT use water, solvents, detergents or spray on electrical parts.
9.1.3 Incubation oven cleaning
Be sure that the autosampler is in stand-by status and that no other operations are in progress.
Set the oven stand-by temperature to off (see paragraph 7.9.2 “Eco Savings”);
Enter “manual operations and open the oven cover (see paragraph 4.2.5 “Setup: Manual
operations”);
Turn OFF the autosampler and remove the power cable;
Wait until the incubator oven cools down;
Remove the spacers (if present) using the suitable tweezers ;
Clean the incubator and the spacers;
Reinstall the spacers (if needed);
Let the oven dry;
Connect the autosampler to the power supply and switch ON the unit;
Warning
Oven surface and spacers (when inserted) are heated at high temperature and can be
very hot. We recommend taking care in order to prevent burns.