HT2800T_en_c.doc Page 85 of 275
•
Rear Injector Insertion Speed: shows the needle insertion speed into the injector.
This parameter is shown only if the autosampler is set to work with the rear injector;
•
IP Address, Subnet Mask, Gateway: shows the LAN connection data (see
paragraph 3.5.1.1 “Default IP address“).
Select “Exit” to exit from the “VIEW PARAMETERS” screen.
4.2 Setup
Access sequence: RUN>SETTINGS>SETUP
In the “SETUP” screen the following menus are available:
•
Alignment: aligns injectors and edit rack settings (see paragraph 4.2.1 “Setup:
Alignment”);
•
Syringe Installation: used to install the syringe (see paragraph 4.2.2 “Setup:
Syringe warmer assembly installation/ replacement (Headspace mode)” for
Headspace mode and paragraph 4.2.3 “Syringe installation/ replacement (liquid mode)“)
for liquid mode;
•
Fiber installation: used to install the Fiber (see paragraph 4.2.4 “SPME Fiber holder
installation/ replacement“). Only in SPME mode;
•
Manual Operation: moves the autosampler motors for diagnostic or emergency needs
(see paragraph 4.2.5 “Setup: Manual operations“);
•
Configuration: edits autosampler settings (see paragraph 4.2.6 “Setup: Configuration”);
•
Service: maintenance operation on the autosampler (see paragraph 0 “Select “Exit” to
exit from this screen and return to the “Setup Configuration”” screen (see
paragraph 4.2.6 “Setup: Configuration”):
•
Setup: Service“).
Select “Exit” to exit from the “SETUP” screen.
4.2.1 Setup: Alignment
This is used to align the sample tray and GC injection port.
Access sequence: RUN>SETTINGS>SETUP>ALIGNMENT
From this screen access the following sub menu:
•
Analyzer Default: selects the correct analyzer on which the autosampler is mounted
(see paragraph (see paragraph 4.2.1.1 “Alignment: Analyzer Default“). If no analyzer has
been selected, the front and rear injector buttons will not be available;