HT2800T_en_c.doc Page 86 of 275
Front injector: aligns the front injector (see paragraph 4.2.1.2 “Alignment: Front
Injector“);
Rear injector: aligns the rear injector (see paragraph 4.2.1.3 “Alignment: Rear Injector“);
Touch Plunger Zero/Touch Fiber Zero: to re-execute alignment procedure (see paragraph 0
“4.2.1 “Setup: Alignment”).
Alignment: Touch Plunger Zero (Headspace and Liquid mode)/ Touch Fiber Zero (SPME
mode)”);
Tray Install: to install and edit rack settings (see paragraph 4. 2. 1. 5 Alignment:
Tray Install “).
Select “Exit to exit from this screen and return to the “Setup” screen (see paragraph
4. 24.2 “Setup“).
4.2.1.1Alignment: Analyzer Default
Access sequence: RUN>SETTINGS>SETUP>ALIGNMENT>ANALYZER DEFAULT
The first screen (“Setup: Analyzer Brand”) shows a list of analyzer brands. Each brand is
associated with a list of models (second screen: “Set up: Analyzer Model”).
Verify that the selected analyzer (marked blue) matches your analyzer.
If not, select the brand of your analyzer and then the model. If your analyzer is not present in the
list select “Generic” and then set the number of injection ports.
Once complete, any previous data relating to the installation will be erased.
Select Exit to exit from this screen and return to the “ALIGNMENTscreen (see paragraph
4.2.1 “Setup: Alignment“). Perform the alignment on the front injector (see paragraph 4.2.1.2
“Alignment: Front Injector“) and rear injector if present (see paragraph 4.2.1.3 “Alignment: Rear
Injector“).
4.2.1.2 Alignment: Front Injector
Access sequence: RUN>SETTINGS>SETUP>ALIGNMENT>FRONT INJECTOR
When “Front injector” is selected, the screen shows Wait: Front Injector Alignment in progress””
(Tap “STOP”, if you want to end this procedure). The autosampler will go directly to the last
position stored in memory or to the default position for the selected analyzer (4.2.1.1 “Alignment:
Analyzer Default“).
-If the vial locator finds the injector, the following message appears: “Confirm Front injector
Alignment?”
Tap “CONFIRM” if the autosampler is correctly aligned over the injector. Set the Front