1.6 Functions available only through HTA Autosampler
Manager
1.6.1 Vial Leakage Check (Headspace Mode)
HTA Autosampler Manager allows the user to enable the Vial Leakage Check (ACCESS
SEQUENCE: Setup-Options).
If this option is enabled, the pressure inside vials is monitored by an heuristic procedure in order
to check against anomalous values that are indicative of a vial leakage problem.
The Vial leakage test can discriminate for samples which are correctly sealed vs. bad crimping
or missing septa. Each sample is marked with PASS or FAIL information in HTA Autosampler
Manager Windows.
1.6.2 Method development option (Headspace Mode)
During method development it can be useful to test methods characterized by a progressive
increment of a specific parameter in order to define which are the best settings for this
parameter.
The parameters that can be increased in this way are the syringe/oven temperature and the
incubation time.
This option is only available with “HTA Autosampler Manager”, using the “Method development”
tab. For the detailed explanation of this option, please refer to “HTA Autosampler Manager”
online help.
1.7 Versions
1.7.1 Description
The autosampler is available in the following versions:
HT2800T
This version is equipped with a touch screen interface.
It includes Headspace and Liquid kit (SPME can be sourced separately, also at a later time).
HT2850T