EasyManua.ls Logo

HTA HT2800T - Page 194

HTA HT2800T
275 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HT2800T_en_c.doc Page 194 of 275
Figure 167: Mount Headspace vial locator
4) “Select syringe volume” screen: select the volume of the installed syringe and press
“SAVE”.
5) “Insert new rack” screen: remove the rack for 2ml vial and install the rack for 6/10/20ml
vials, then press “CONTINUE”.
Figure 168: Rack for 6/10720ml vials installation
6) “Remove spacers and insert vials in position A4, C and System Integrity Tool in proper
positionscreen: remove the six black spacers for liquid configuration in the oven using
the supplied tweezers (see paragraph 2.3 Tool kit“), add a headspace sample vial
12
in
A.4 position of the rack and in position C of the oven and a the System Integrity Tool
(only if you purchased the GLP pack, see paragraph (1.7.2 “Options”). into the dedicated
12
The headspace vial to be put in the rack must have the same volume of the vial used last time in
headspace mode.

Table of Contents

Related product manuals