HT2800T_en_c.doc Page 129 of 275
Desorption
Synchro
The autosampler starts
sample preparation
The autosampler starts
the desoprtion
The autosampler
gives the “Start”
signal to the
analyzer
A-Start
At the reception of the
“Ready” signal from the
analyzer
After ending the sample
preparation the autosampler
checks again for the
presence of the “Ready”
signal from the analyzer and
then starts the desorption
When the autosampler
touches the injector
D-Start
At the reception of the
“Ready” signal from the
analyzer
After ending the sample
preparation the autosampler
checks again for the
presence of the “Ready”
signal from the analyzer and
then starts the desorption
At the end of the
syringe plunger
movement
Normal
w/o Rdy
At the end of the analysis
time set in the autosampler
method
After the ending sample
preparation
At the beginning of the
syringe plunger
movement
Clean mode: only for SPME mode. It sets the predefined clean mode. In HT2800T this
parameter is editable only if the system is configured for two injectors. This setting is
used for single injection runs and automatic run.
Clean Mode
Front: the Fiber cleaning is performed in the first injector
Rear: the Fiber cleaning is performed in the second injector
Cleaning Device: the Fiber cleaning is performed in the SPME Fiber
Cleaning Devive location (indicated with number 7 in Figure 1). This option
is available only in HT2850T.
Shaker speed. It sets the speed of the shaking (only in Headspace and SPME mode).
Empty Position Synchro. It sets the autosampler behaviour when a vial is missing (only
in Liquid mode). It may have the following values:
Empty
Position
Synchro
Yes: If a vial is missing, the autosampler emits the “Start” signal to the
analyzer
“as
if
the vial
were
present”.
No: If a vial is missing, the autosampler does not emit the “Start” signal to
the analyzer and skips directly to the next vial.