Follow this procedure to move from Liquid to Headspace mode Access sequence: RUN>CHANGE
MODE>Headspace
The screen “You are moving to Headspace modeconfirms the start of the changing procedure.
Tap “CONTINUEto start the switching operation or ABORT” to return to the “CHANGE MODE”
screen.
Open the syringe location by pushing up the sliding lid and follow the wizard:
1) “Remove vial locator” screen: remove the liquid vial locator using the supplied allen key
to loose the fixing screws ( see paragraph 2.3 Tool kit“) and tap “CONTINUE”.
Figure 161: Change from liquid to Headspace mode_Remove vial locator
2) “Remove and install new syringe” screen:
Warning
If syringeID system is used, the syringe information are loaded in a RFID tag included in
the black adapter mounted on the syringe for the syringeID system (see paragraph 11.2
“Liquid mode“). Every adapter must be associated to a syringe only in order to maintain
the info regarding the syringe always correct and updated. Do not remove the syringe
from its black adapter.