HT2800T_en_c.doc Page 50 of 275
Installation operations should be performed using the touch screen display.
1. Install the autosampler on the mounting plate (see paragraph 3.3.1 Autosampler
installation on the mounting plate” ).
2. Move the safety lock from the shipping position to the operating position (see paragraph
3.3.2 “Safety Lock setting”).
3. Connect cables (see paragraph 3.3.3 “Electrical connections”).
4. Connect purge line to be used only in Headspace and for Fiber Cleaning Device (see
paragraph 3.3.4 “Purge line connections (for Headspace and Fiber Cleaning Device)“).
5. Switch on the autosampler (see paragraph 3.4 “Start-Up “).
6. Install PC software (see paragraph 3.5 “PC connection and software installation“) (only if
it has been purchased).
7. Install the Headspace syringe. See paragraph 4.2.2 Setup: Syringe warmer
assembly installation/ replacement (Headspace mode).
8. Select the analyzer (see paragraph 4.2.1.1 Alignment: Analyzer Default”).
9. Align the autosampler on the GC Injector(s) (see paragraph 4.2.1.2 Alignment:
Front Injector” and Errore. L'origine riferimento non è stata trovata.Errore. L'origine
riferimento non è stata trovata.”.)
10. Install the sample rack for 6/10/20ml vials (see paragraph 2 .1. 1.1
HT2800T/HT2850T Sample rack,paragraph 7.7 “Load and Unload Sample Vial” to open
the tray and paragraph 7.9.1.1 “Rack mounting” to correctly mount the rack).
Perform the touch and plunger zero procedure (see paragraph 0 4.2.1 “Setup: Alignment”).
11. Alignment: Touch Plunger Zero (Headspace and Liquid mode)/ Touch Fiber Zero (SPME
mode)“).
12. If SPME kit (see paragraph 1 . 7 . 2 Options “) has been purchased go to point
13), if SPME has not been purchased go to point 15.
13. Change from headspace to SPME mode (see paragraph 7.10.5 Change from
Headspace to SPME mode “). At the end of the change mode procedure, it is always
necessary to adjust the injector alignment (see point below).
14. Align the autosampler on the GC Injector(s) (see paragraphs 4.2.1.2 Alignment:
Front Injector” and Errore. L'origine riferimento non è stata trovata.Errore. L'origine
riferimento non è stata trovata.”).
15. Change to liquid mode (see paragraph 7.10.2 Change from SPME to Liquid mode “ or