HT2800T_en_c.doc Page 91 of 275
skipped by tapping the appropriate button. Select “Exit” to exit from this screen and return to the
“ALIGNMENT” screen (see paragraph (see paragraph 4.2.1 “Setup: Alignment”).
4.2.1.5 Alignment: Tray Install
Access sequence: RUN>SETTINGS>SETUP>ALIGNMENT>TRAY INSTALL
This screen shows a list of trays/racks stored in memory.
The following options are available:
•
Add new: adds a new tray/rack;
•
Remove: removes a tray/rack;
Select a tray to view or edit its details. The number of vial positions, dimension of the sample
vials and the needle draw depth inside the sample vial can be modified if your sampler allows it.
If you modify a parameter, a screen displays: “Modifying tray”. Tap “SAVE” to confirm, otherwise
tap “CANCEL”. A confirmation screen appears “Tray parameters has been modified”. Tap
“CONTINUE“.
Select “ADD NEW”, to set the number of vial positions, the dimension of the sample vials and
the needle draw depth inside the sample vial of the new tray/rack (e.g.: 121-2ml-30mm).
To remove a rack, select “REMOVE” then select the tray/rack that you want to remove. A
confirmation screen appears “The selected tray has been removed”. Tap “CONTINUE”.
Select “Exit” to exit from this screen and return to the “ALIGNMENT” screen (see paragraph
4.2.1 “Setup: Alignment”).
Warning
The suggested needle draw depth inside the sample vial for Liquid mode is 30mm if
12x32mm (2ml) vial without micro-inserts are used (12.2.2 “Sample vials“). 32mm is the
maximum depth.