EasyManuals Logo

Keithley 2010 User Manual

Keithley 2010
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #127 background imageLoading...
Page #127 background image
*ESE <NRf> Event Enable Program the standard event enable register
*ESE? Event Enable Query Read the standard event register
Parameters
<NRf> = 0 Clear register
1 Set OPC (B0)
4 Set QYE (B2)
8 Set DDE (B3)
16 Set EXE (B4)
32 Set CME (B5)
64 Set URQ (B6)
128 Set PON (B7)
255 Set all bits
Description
Use the *ESE command to program the Standard Event Enable Register. This command is
sent with the decimal equivalent of the binary value that determines the desired state (0 or 1) of
the bits in the register. This register is cleared on power-up.
This register is used as a mask for the Standard Event Register. When a standard event is
masked, the occurrence of that event will not set the Event Summary Bit (ESB) in the Status
Byte Register. Conversely, when a standard event is unmasked (enabled), the occurrence of that
event sets the ESB bit. For information on the Standard Event Register and descriptions of the
standard event bits, see the following section.
A cleared bit (0) in the enabled register prevents (masks) the ESB bit in the Status Byte
Register from setting when the corresponding standard event occurs. A set bit (1) in the enable
register allows (enables) the ESB bit to set when the corresponding standard event occurs.
Remote Operation 4-35

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Resistance Accuracy0.01%
InterfaceGPIB, RS-232
Power Supply50 Hz or 60 Hz