Optimizing measurement accuracy
The configurations listed below assume that the multimeter has had factory setups restored.
DC voltage, DC current, and resistance:
• Select 6½ digits, 10 PLC, filter ON (up to 100 readings), fixed range.
• Use REL on DC voltage and two-wire resistance measurements.
• Use four-wire resistance measurements for best accuracy.
AC voltage and AC current:
• Select 6½ digits, 10 PLC, filter ON (up to 100 readings), fixed range.
Temperature:
• Select 6½ digits, 10 PLC, filter ON (up to 100 readings).
Optimizing measurement speed
The configurations listed below assume that the multimeter has had factory setups restored.
DC voltage, DC current, and resistance:
Select 3½ digits, 0.01 PLC, filter OFF, fixed range.
AC voltage and AC current:
Select 3½ digits, 0.01 PLC, filter OFF, fixed range.
Temperature:
• Select 3½ digits, 0.01 PLC, filter OFF.
For all functions, turn off the display and autozero and set the trigger delay to zero. Use the
:SAMPle:COUNt and READ? bus commands.
Specifications A-11