:IMMediate
:CALCulate3:IMMediate Perform CALC3
Description When the configuration of the limit test is changed, the next reading is
evaluated according to the new test configuration. If the instrument is not in
a continuous measurement mode (e.g., waiting for a manual trigger), the test
will not be performed until the next reading conversion occurs.
This action command lets you re-process the current input data to test new
limits. For example, assume the instrument is in a non-continuous
measurement mode and requires a manual trigger to cause the next reading
conversion. Changing the test limits will not affect the last test result.
However, sending the :IMMediate command reprocesses the data and
evaluates the reading according to the new test limits. Note that sending the
:IMMediate command does not initiate a reading conversion.
Program PRINT #1, “output 16;:trig:sour bus” ‘ Place 2010 in one-shot mode
SLEEP 3 ‘ Wait three seconds
PRINT #1, “output 16;:calc3:imm” ‘ Re-perform limit test
DISPlay subsystem
The commands in this subsystem are used to control the display of the
Model 2010 and are summarized in Table 5-3.
:ENABle <b>
:DISPlay:ENABle <b> Control display circuitry
Parameters <b> = 0 or OFF Disable display circuitry
1 or ON Enable display circuitry
Query :ENABle? Query state of display
Description This command is used to enable and disable the front panel display circuitry.
When disabled, the instrument operates at a higher speed. While disabled, the
display is frozen.
All front panel controls (except LOCAL) are disabled. Normal display
operation can be resumed by using the :ENABle command to enable the
display or by putting the Model 2010 into local mode (press LOCAL).
5-26 SCPI Command Reference