Shielding
To achieve a stable reading, shield resistances greater than 100k. Place the resistance in a
shielded enclosure and connect the shield to the INPUT LO terminal of the instrument
electrically.
See Section 3 for information on the configuration options for two-wire and four-wire
resistance measurements.
Low resistance measurements
The Model 2010 can be used for low resistance measurements normally handled by a micro-
ohmmeter. The following paragraphs discuss the Model 2010’s dry circuit testing and offset
compensation modes.
2001 MULTIMETER
Model 2010
Resistance
Under Test
Shielded
Cable
Optional Shield
Note: Source current flows from the INPUT
HI to INPUT LO terminals.
SHIFT
CH1REM
TALK
LSTN
SRQ
STATREL FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH
8
CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
2001 MULTIMETER
Resistance
Under Test
Shielded
Cable
Optional Shield
Note: Source current flows from the INPUT HI
to INPUT LO terminals.
Model 2010
SHIFT
CH1REM
TALK
LSTN
SRQ
STATREL FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH
8
CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
Fi
gure
2
-
6
Two- and four-wire
resistance
measurements
Basic Measurements 2-25

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
BrandKeithley
Model2010
CategoryMultimeter
LanguageEnglish

Related product manuals