Speed Commands
:NPLCycles <n>
[:SENSe[1]]:CURRent:AC:NPLCycles <n> Set NPLC for ACI
[:SENSe[1]]:CURRent[:DC]:NPLCycles <n> Set NPLC for DCI
[:SENSe[1]]:VOLTage:AC:NPLCycles <n> Set NPLC for ACV
[:SENSe[1]]:VOLTage[:DC]:NPLCycles <n> Set NPLC for DCV
[:SENSe[1]]:RESistance:NPLCycles <n> Set NPLC for Ω2
[:SENSe[1]]:FRESistance:NPLCycles <n> Set NPLC for Ω4
[:SENSe[1]]:TEMPerature:NPLCycles <n> Set NPLC for TEMP
Parameters <n> = 0.01 to 10 Power line cycles per integration
DEFault 1
MINimum 0.01
MAXimum 10
Query :NPLCycles? Query programmed NPLC value
:NPLCycles? DEFault Query *RST default value
:NPLCycles? MINimum Query minimum NPLC value
:NPLCycles? MAXimum Query maximum NPLC value
Description The integration period (measurement speed) for the basic measurement
functions (except Frequency and Period) is set using the :NPLCycle
command. NPLC (Number of Power Line Cycles) expresses the integration
period by basing it on the power line frequency. For example, for a PLC of
1, the integration period in seconds would be 1/60 (for 60Hz line power)
which is 16.67msec.
5-40 SCPI Command Reference