:DATA command
:DATA?
[:SENSe[1]]:DATA? Return reading.
Description This query command is used to read the latest instrument reading. This
command returns the raw reading or a reading that is the result of the
Reference (REL from the front panel) operation. For example, if a reference
value of 1.0 is established, the reading returned by this command is the raw
reading minus 1.0. Calculated (MATH) readings cannot be read with this
command (see the CALCulate subsystem for information on how to read
math calculations).
The reading is returned in exponent form. For example, a 10V DC reading
will be displayed on the CRT as +1.000000E+01.
The measurement function is not included in the response message. Thus,
you may want to perform a function query (see previous command) after a
reading query.
:FRESh?
[:SENSe[1]]:DATA:FRESh? Return new reading.
Description This query command is used to return a new (fresh) reading. This command
will not request the same reading twice. If a new reading is triggered, this
command will wait until the reading is available, rather than request the old
reading.
5-38 SCPI Command Reference

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Resistance Accuracy0.01%
InterfaceGPIB, RS-232
Power Supply50 Hz or 60 Hz

Related product manuals