:ALPHa <NRf>
[:SENSe[1]]:TEMPerature:FRTD:ALPHa <NRf> Specify alpha value
Parameters <NRf> = 0 to 0.01 Specify FRTD alpha value
Query :ALPHa? Query value of ALPHa
Description This command is used to check and/or change the alpha value. Remember
that if you change the present alpha value, the FRTD TYPE will change to
USER.
:BETA <NRf>
[:SENSe[1]]:TEMPerature:FRTD:BETA <NRf> Specify beta value
Parameters <NRf> = 0 to 1.00 Specify FRTD beta value
Query :BETA? Query value of BETA
Description This command is used to check and/or change the beta value. Remember that
if you change the present beta value, the FRTD TYPE will change to USER.
:DELTa <NRf>
[:SENSe[1]]:TEMPerature:FRTD:DELTa <NRf> Specify delta value
Parameters <NRf> = 0 to 5.00 Specify FRTD delta value
Query :DELTa? Query value of DELTa
Description This command is used to check and/or change the delta value. Remember
that if you change the present delta value, the FRTD TYPE will change to
USER.
:DIODe command
:RANGe[:UPPer] <NRf>
[:SENSe[1]]:DIODe:CURRent:RANGe[:UPPer] <NRf> Select current range for diode test
Parameters <NRf> = 0 to 1e-3 Specify diode test current
Query [UPPer]? Query selected range
Description There are three current ranges available for the diode test: 10µA range,
100µA range, and the 1mA range. Range is selected by using this command
to specify the expected current for the diode under test. The instrument will
then automatically select the appropriate range.
SCPI Command Reference 5-53