5-64 SCPI Command Reference
:SYSTem subsystem
The SYSTem subsystem contains miscellaneous commands that are
summarized in Table 5-8.
:BEEPer command
:STATe <b>
:SYSTem:BEEPer:STATe <b> Enable or disable beeper
Parameters <b> = 1 or ON Enable beeper
0 or OFF Disable beeper
Query :STATe? Query state of beeper
Description This command is used to enable or disable the beeper for limit tests.
:PRESet command
:PRESet
:SYSTem:PRESet Return to :SYSTem:PRESet defaults
Description This command returns the instrument to states optimized for front panel
operation. :SYSTem:PRESet defaults are listed in the SCPI tables
(Table 5-2 through Table 5-11).
:KCLick command
:KCLick <b>
:SYSTem:KCLick <b> Enable or disable keyclick
Parameters <b> = 1 or ON Enable keyclick
0 or OFF Disable keyclick
Query :KCLick? Query status of keyclick
Description This command is used to enable or disable the keyclick. The keyclick can
also be enabled or disabled from the front panel by pressing SHIFT then
LOCAL.

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Resistance Accuracy0.01%
InterfaceGPIB, RS-232
Power Supply50 Hz or 60 Hz