Measuring temperature
The Model 2010 can measure temperature with a four-wire RTD sensor or a thermocouple.
The temperature measurement ranges available depend on the type of RTD or thermocouple
chosen.
RTDs can be connected to the banana jacks on the front or rear panel.
Thermocouples can be connected to the Model 2001-TCSCAN card, which plugs into the
option slot of the Model 2010, or to an external thermocouple card, such as a Model 7057A,
7402, or 7014 installed in a Model 7001 or 7002 Switch System.
If the Model 2001-TCSCAN card is not used then an estimate of the panel temperature must
be made (usually 2°C above room temperature). Connect the thermocouple card directly to the
front panel Input HI and LO as shown in Figure 2-9. To input the panel temperature estimate,
choose the TCOUPLE configuration option and select JUNC. Pick the SIM option and input the
estimate.
2-30 Basic Measurements

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Resistance Accuracy0.01%
InterfaceGPIB, RS-232
Power Supply50 Hz or 60 Hz