EasyManua.ls Logo

Keithley 2010 - Threshold Commands; Transducer Commands

Keithley 2010
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:THReshold commands
Use these commands to set the maximum range input (signal level) for
frequency and period measurements.
:RANGe <n>
[:SENSe[1]]:PERiod:THReshold:VOLTage:RANGe <n> Set voltage threshold range
[:SENSe[1]]:FREQuency:THReshold:VOLTage:RANGe <n> Set voltage threshold range
Parameters <n> = 0 to 1010 Specify signal level in volts (voltage threshold)
Query :RANGe? Query maximum signal level
Description These commands are used to specify the expected input level. The instrument
will then automatically select the most sensitive current or voltage threshold
range.
:TRANsducer commands
:TRANsducer <name>
[:SENSe[1]]:TEMPerature:TRANsducer <name> Select sensor type
Parameters <name> = FRTD Set sensor type to four-wire RTD
TCouple Set sensor type to thermocouple
Query :TRANsducer? Query type of temperature sensor
Description This command is used to specify the type of sensor desired for temperature
measurements. Specify TCouple if you are using a thermocouple. The
:TCouple:TYPE command is then used to specify which type of
thermocouple you wish to use. Specify FRTD if you are using a four-wire
RTD sensor. The :FRTD:TYPE command is then used to specify a particular
four-wire RTD sensor type.
SCPI Command Reference 5-49

Table of Contents

Other manuals for Keithley 2010

Related product manuals