:THReshold commands
Use these commands to set the maximum range input (signal level) for
frequency and period measurements.
:RANGe <n>
[:SENSe[1]]:PERiod:THReshold:VOLTage:RANGe <n> Set voltage threshold range
[:SENSe[1]]:FREQuency:THReshold:VOLTage:RANGe <n> Set voltage threshold range
Parameters <n> = 0 to 1010 Specify signal level in volts (voltage threshold)
Query :RANGe? Query maximum signal level
Description These commands are used to specify the expected input level. The instrument
will then automatically select the most sensitive current or voltage threshold
range.
:TRANsducer commands
:TRANsducer <name>
[:SENSe[1]]:TEMPerature:TRANsducer <name> Select sensor type
Parameters <name> = FRTD Set sensor type to four-wire RTD
TCouple Set sensor type to thermocouple
Query :TRANsducer? Query type of temperature sensor
Description This command is used to specify the type of sensor desired for temperature
measurements. Specify TCouple if you are using a thermocouple. The
:TCouple:TYPE command is then used to specify which type of
thermocouple you wish to use. Specify FRTD if you are using a four-wire
RTD sensor. The :FRTD:TYPE command is then used to specify a particular
four-wire RTD sensor type.
SCPI Command Reference 5-49

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
BrandKeithley
Model2010
CategoryMultimeter
LanguageEnglish

Related product manuals