SCPI Command Reference 5-65
:POSetup <name> command
:POSetup <name>
:SYSTem:POSetup <name> Program power-on defaults
Parameters <name> = RST Select *RST defaults on power up
PRESet Select :SYSTem:PRESet defaults on power up
SAV0 Select saved defaults on power up
Query :POSetup? Query power-on setup
Description This command is used to select the power-on defaults. With RST selected,
the instrument powers up to the *RST default conditions. With PRES
selected, the instrument powers up to the :SYStem:PRESet default
conditions. Default conditions are listed in the SCPI tables (Table 5-2
through Table 5-11).
With the SAV0 parameter selected, the instrument powers-on to the setup
that is saved in the specified location using the *SAV command.
:FRSWitch?
:SYSTem:FRSWitch? Read INPUTS switch
Description This query command is used to read the position of the FRONT/REAR
INPUTS switch. Switch position code is defined as follows:
1 = Front panel inputs selected
0 = Rear panel inputs selected
:VERSion? command
:VERSion?
:SYSTem:VERSion? Read SCPI version
Description This query command is used to read the version of the SCPI standard being
used by the Model 2010. Example code:
1991.0
The above response message indicates the version of the SCPI standard.