SCPI Command Reference 5-73
Trigger subsystem
The Trigger subsystem is made up of a series of commands and subsystems
to configure the Trigger Model. These commands and subsystems are
summarized in Table 5-10.
:INITiate commands
[:IMMediate]
:INITiate[:IMMediate] Take 2010 out of idle state
Description This command takes the Model 2010 out of the idle state. After all
programmed operations are completed, the instrument returns to the idle
state if continuous initiation is disabled (see next command).
:CONTinuous <b>
:INITiate:CONTinuous <b> Control continuous initiation
Parameters <b> = 0 or OFF Disable continuous initiation
1 or ON Enable continuous initiation
Query :CONTinuous? Query continuous initiation
Description When continuous initiation is selected (ON), the instrument is taken out of
the idle state. At the conclusion of all programmed operations, the instrument
returns to the top of the trigger model.
NOTE: With continuous initiation enabled (ON), you cannot use the
:READ? command or set sample count greater than one (see
:SAMPle:COUNt).
:ABORt command
:ABORt Abort operation
Description When this action command is sent, the Model 2010 aborts operation and
returns to the top of the Trigger Model. If continuous initiation is disabled,
the instrument goes to the idle state. If continuous initiation is enabled,
operation continues at the top of the trigger model.
The abort command resets the scan pointer back to the first channel in the
scan list.

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Resistance Accuracy0.01%
InterfaceGPIB, RS-232
Power Supply50 Hz or 60 Hz

Related product manuals