:COUNt <n>
[:SENSe[1]]:CURRent:AC:AVERage:COUNt <n> Specify lter count for ACI
[:SENSe[1]]:CURRent[:DC]:AVERage:COUNt <n> Specify lter for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:COUNt <n> Specify lter count for ACV
[:SENSe[1]]:VOLTage[:DC]:AVERage:COUNt <n> Specify lter count for DCV
[:SENSe[1]]:RESistance:AVERage:COUNt <n> Specify lter count for 2
[:SENSe[1]]:FRESistance:AVERage:COUNt <n> Specify lter count for 4
[:SENSe[1]]:TEMPerature:AVERage:COUNt <n> Specify lter count for TEMP
Parameters <n> = 1 to 100 Specify filter count
DEFault 10
MINimum 1
MAXimum 100
Query :COUNt? Query filter count
:COUNt? DEFault Query the *RST default filter count
:COUNt? MINimum Query the lowest allowable filter count
:COUNt? MAXimum Query the largest allowable filter count
Description These commands are used to specify the filter count. In general, the filter
count is the number of readings that are acquired and stored in the filter
buffer for the averaging calculation. The larger the filter count, the more
filtering performed.
Bandwidth command
:BANDwidth <n>
[:SENSe[1]]:CURRent:AC:DETector:BANDwidth <n> Specify maximum bandwidth for ACI
[:SENSe[1]]:VOLTage:AC:DETector:BANDwidth <n> Specify maximum bandwidth for ACV
Parameters <n> = 3 to 300e3 Specify bandwidth (in Hz)
Query BANDwidth? Query selected bandwidth
Description The Model 2010 uses three bandwidth settings for ACI and ACV
measurements; 3 (3Hz-300kHz), 30 (30Hz-300kHz), and 300 (300Hz-
300kHz). To achieve the best accuracy, you should use the bandwidth setting
that best reflects the frequency of the input signal. For example, if the input
signal is 40Hz, then a bandwidth setting of 30 should be used.
These commands are used to select bandwidth for the ACI and ACV
functions. To set the bandwidth, simply specify (approximately) the
frequency of the input signal. The Model 2010 will automatically select the
optimum bandwidth setting.
NOTE For bandwidth setting of 3 and 30, the normal A/D reading
conversion method is not used. Thus, the NPLC setting is only
valid for bandwidth setting of 300.
5-48 SCPI Command Reference

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
BrandKeithley
Model2010
CategoryMultimeter
LanguageEnglish

Related product manuals