Program Fragment
PRINT #1, "output 16; :syst:pres" 'Select defaults
PRINT #1, "output 16; :init:cont off;:abort"
'Place 2010 in idle
PRINT #1, "output 16; :trig:coun 1; sour tim"
PRINT #1, "output 16; :samp:coun 5" 'Program for five measurements
'and stop (idle)
PRINT #1, "output 16; :init; *opc?" 'Start measurements and send
'*opc?
PRINT #1, "enter 16" 'Get response when 2010 goes into
'idle
LINE INPUT #2, a$ 'Read contents of Output Queue
PRINT a$ 'Display the ASCII "1"
*OPT? Option Identication Query Determine if an option is installed
Description
The response message indicates the presence or absence of an optional scanner card. For
example:
0 No scanner card installed
200X-SCAN Scanner card installed
*RCL Recall Return to setup stored in memory
Parameters
<NRf>=0
Description
Use this command to return the Model 2010 to the configuration stored in memory. The *SAV
command is used to store the setup configuration in memory location.
Only one setup configuration can be saved and recalled.
The Model 2010 ships from the factory with :SYSTen:PRESet defaults loaded into the
available setup memory. If a recall error occurs, the setup memory defaults to the
:SYSTem:PRESet values.
Remote Operation 4-41

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
BrandKeithley
Model2010
CategoryMultimeter
LanguageEnglish

Related product manuals