:OPEN <list>
:ROUTe:MULTiple:OPEN <list> Open specied channels
Parameter <list> = (@ chanlist) Specify channels to open
where chanlist is the list of channels (1 through 11) to be opened.
Description This command is used to open specified channels on the internal scanner
card. When this command is sent, the channels specified in the chanlist will
open.
Channel 11 is the 2-pole/4-pole relay. Opening channel 11 selects the 4-pole
operating mode.
Examples of a list:
list = (@1,3,5) Channels 1, 3, and 5.
= (@1:5) Channels 1 through 5.
:SCAN commands
[:INTernal] <list>
:ROUTe:SCAN[:INTernal] <list> Dene internal scan list and enable scan.
Parameter <list> = (@ scanlist)
where scanlist is the specified list of channels (1 through 10) to be scanned.
Query [:INTernal]? Query programmed scan list
Description This command is used to define the scan list for the internal scanner card. The
scan list can contain two to ten channels. The following examples
demonstrate the various forms for expressing a scan list:
list = (@ 2,3,4) Channels separated by commas (,).
(@ 1:8) Range of channels (1 through 8). Range limits separated
by a colon (:).
NOTE You can only scan consecutive channels. Skipping channels is not allowed. For
example:
(@1:4) is valid.
(@1,2,4) is not valid.
See the instruction manual of the internal scanner card for details on scanning.
SCPI Command Reference 5-35

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
BrandKeithley
Model2010
CategoryMultimeter
LanguageEnglish

Related product manuals