:EXTernal <list>
:ROUTe:SCAN:EXTernal <list> Define external scan list
Parameter <list> = (@ scanlist)
where scanlist is the specified list of external channels (1 to 800) to be
scanned.
Query :EXTernal? Query programmed scan list
Description The Model 2010 can operate with an external switch system, such as the
Keithley Model 7001 or 7002. The Model 2010 can measure up to 800
channels that are switched by the external switching system. This command
is used to define the external scan list.
The scan list can contain 2 to 800 channels. See :SCAN[:INTernal] for
examples to express a scan list. The external scan is enabled by the
ROUTe:SCAN:LSELect EXTernal command.
:LSELect <name>
:ROUTe:SCAN:LSELect <name> Perform specified scan operation
Parameters <name> = INTernal Enable scan for internal scanner card
EXTernal Enable scan for external scanner card
NONE Disable all scan operations
Query :LSELect? Query scan operation
Description This command is used to select and perform the desired scan operation.
When INTernal is selected, the Model 2010 scans the channels of the internal
switching card according to how the scan is configured (see
:SCAN[:INTernal]).
EXTernal is used to measure channels that are controlled by an external
switch system. When EXTernal is selected, the Model 2010 scans the
external scan list (see :SCAN:EXTernal).
When NONE is used, the Model 2010 disables all operations associated with
a scan.
5-36 SCPI Command Reference