:AVERage commands
The :AVERage commands are used to configure and control the filter. The
Filter is explained in Section 3.
:STATe <b>
[:SENSe[1]]:CURRent:AC:AVERage:STATe <b> Control lter for ACI
[:SENSe[1]]:CURRent[:DC]:AVERage:STATe <b> Control lter for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:STATe <b> Control lter for ACV
[:SENSe[1]]:VOLTage[:DC]:AVERage:STATe <b> Control lter for DCV
[:SENSe[1]]:RESistance:AVERage:STATe <b> Control lter for 2
[:SENSe[1]]:FRESistance:AVERage:STATe <b> Control lter for 4
[:SENSe[1]]:TEMPerature:AVERage:STATe <b> Control lter for TEMP
Parameters <b> = 0 or OFF Disable the digital filter
1 or ON Enable the digital filter
Query :STATe? Query state of digital filter
Description These commands are used to enable or disable the digital averaging filter for
the specified function. When enabled, readings will be filtered according to
how the filter is configured.
:TCONtrol <name>
[:SENSe[1]]:CURRent:AC:AVERage:TCONtrol <name> Select lter type for ACI
[:SENSe[1]]:CURRent[:DC]:AVERage:TCONtrol <name> Select lter type for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:TCONtrol <name> Select lter type for ACV
[:SENSe[1]]:VOLTage[:DC]:AVERage:TCONtrol <name> Select lter type for DCV
[:SENSe[1]]:RESistance:AVERage:TCONtrol <name> Select lter type for 2
[:SENSe[1]]:FRESistance:AVERage:TCONtrol <name> Select lter type for 4
[:SENSe[1]]:TEMPerature:AVERage:TCONtrol <name> Select lter type for TEMP
Parameters <name> = REPeat Select repeating filter
MOVing Select moving filter
Query :TCONtrol? Query filter type
Description These commands are used to select the type of averaging filter (REPeat or
MOVing) for the specified function. These filter types are explained in
Section 3.
The number of readings that are averaged by the filter is set with the
:AVERage:COUNt command. The :AVERage:STATe command is used to
enable or disable the filter. Changing the filter type disables auto filter.
SCPI Command Reference 5-47

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
BrandKeithley
Model2010
CategoryMultimeter
LanguageEnglish

Related product manuals