*RST RESET Return 2010 to *RST defaults
Description
When the *RST command is sent, the Model 2010 performs the following operations:
1. Returns the Model 2010 to the *RST default conditions (see SCPI tables).
2. Cancels all pending commands.
3. Cancels response to any previously received *OPC and *OPC? commands.
*SAV Save Save present setup in memory
Parameters
<NRf>=0
Description
Use the *SAVE command to save the present instrument setup configuration in memory for
later recall. Any control affected by *RST can be saved by the *SAV command. The *RCL
command is used to restore the instrument to the saved setup configuration.
Only one setup configuration can be saved and recalled.
*SRE <NRf> Service Request Enable Program service request enable register
*SRE? Service Request Enable Query Read service request enable register
Parameters
<NRf>= 0 Clears enable register
1 Set MSB bit (Bit 0)
4 Set EAV bit (Bit 2)
8 Set QSB bit (Bit 3)
16 Set MAV bit (Bit 4)
32 Set ESB (Bit 5)
128 Set OSB (Bit 7)
255 Set all bits
Description
Use the *SRE command to program the Service Request Enable Register. Send this command
with the decimal equivalent of the binary value that determines the desired state (0 or 1) of each
bit in the register. This register is cleared on power-up.
This enable register is used along with the Status Byte Register to generate service requests
(SRQ). With a bit in the Service Request Enable Register set, an SRQ occurs when the
4-42 Remote Operation

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Resistance Accuracy0.01%
InterfaceGPIB, RS-232
Power Supply50 Hz or 60 Hz