:DIGits command
:DIGits <n>
[:SENSe[1]]:CURRent:AC:DIGits <n> Specify resolution for ACI
[:SENSe[1]]:CURRent[:DC]:DIGits <n> Specify resolution for DCI
[:SENSe[1]]:VOLTage:AC:DIGits <n> Specify resolution for ACV
[:SENSe[1]]:VOLTage[:DC]:DIGits <n> Specify resolution for DCV
[:SENSe[1]]:RESistance:DIGits <n> Specify resolution for Ω2
[:SENSe[1]]:FRESistance:DIGits <n> Specify resolution for Ω4
[:SENSe[1]]:PERiod:DIGits <n> Specify resolution for PER
[:SENSe[1]]:FREQuency:DIGits <n> Specify resolution for FREQ
[:SENSe[1]]:TEMPerature:DIGits <n> Specify resolution for TEMP
Parameters <n> = 4 3½ digits
54½ digits
65½ digits
76½ digits
87½ digits
DEFault 5½ digits for ACI and ACV
6½ digits for TEMP, FREQ, PER
7½ digits for DCI, DCV, Ω2, Ω4
MINimum 3½ digits for ACI, ACV, DCI, DCV, Ω2, Ω4, TEMP,
FREQ, PER
MAXIMUM 6½ digits for ACI, ACV, TEMP, FREQ, PER, Ω4 if
dryckt is on
7½ digits for DCI, DCV, Ω2, Ω4
Query :DIGits? Query selected resolution
:DIGits? DEFault Query *RST default resolution
:DIGits? MINimum Query minimum allowable resolution
:DIGits? MAXimum Query maximum allowable resolution
Description These commands are used to select display resolution for the specified
measurement function.
Even though the parameters for this command are expressed as integers (4 to
7), you can specify resolution using real numbers. For example, to select
3½ digit resolution let <n> = 3.5, for 4½ digit let <n> = 4.5, and so on.
Internally, the instrument rounds the entered parameter value to the nearest
integer.
5-46 SCPI Command Reference