EasyManuals Logo

Keithley 2010 User Manual

Keithley 2010
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #165 background imageLoading...
Page #165 background image
:TEXT commands
:DATA <a>
:DISPlay[:WINDow[1]]:TEXT:DATA <a> Dene message for display.
Parameter <a> = ASCII characters for the message (maximum of 12 characters). The
characters must be enclosed in either double quotes (“ ”) or single
quotes (‘ ’).
Query :DATA? Query the defined text message.
Description These commands define the text message for display. A message can be as
long as 12 characters. A space counts as a character. Excess message
characters results in an error.
:STATe <b>
:DISPlay[WINDow[1]]:TEXT:STATe <b> Control (on/off) message
Parameters <b> = 0 or OFF Disable text message
1 or ON Enable text message
Query :STATe? Query state of message mode.
Description This command enables and disables the text message mode. When enabled,
a defined message is displayed. When disabled, the message is removed from
the display.
A user defined text message remains displayed only as long as the instrument
is in remote. Taking the instrument out of remote (by pressing the LOCAL
key or sending LOCAL 16), cancels the message and disables the text
message mode.
SCPI Command Reference 5-27

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Resistance Accuracy0.01%
InterfaceGPIB, RS-232
Power Supply50 Hz or 60 Hz