Description This command is used to specify the simulated reference temperature. The
temperature value depends on which temperature scale is presently selected
(°C, °F, or K). Typically, 0° or 23°C is used as the simulated reference
temperature.
:REAL:TCOefficient <n>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]]:REAL:TCOefficient <n>
Parameters <n> = -0.09999 to 0.09999 Specify temperature coefficient
DEFault +0.00020 temperature coefficient
MINimum -0.09999 temperature coefficient
MAXimum +0.09999 temperature coefficient
Query :TCOefficient? Query temperature coefficient (TC)
:TCOefficient? DEFault Query *RST default TC
:TCOefficient? MINimum Query lowest allowable TC
:TCOefficient? MAXimum Query largest allowable TC
Description This command is used to specify the temperature coefficient (TC) of the
“real” temperature reference junction. TC is specified in °C/volt and is not
affected by the :UNIT:TEMPerature command.
:REAL:OFFSet <n>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]:REAL:OFFSet <n>
Parameters <n> = -0.09999 to 0.09999 Specify voltage offset at 0°C
DEFault 0.05463
MINimum -0.09999
MAXimum 0.09999
Query :OFFSet? Query voltage offset
:OFFSet? DEFault Query *RST default voltage offset
:OFFSet? MINimum Query lowest allowable voltage offset
:OFFSet? MAXimum Query largest allowable voltage offset
Description This command is used to specify the offset voltage at 0°C for the specified
reference junction.