Specications
Description Data
Transmission rate 10 Mbit/s and 100 Mbit/s
Auto detection (100, FullDX)
Protocol ProNet I/O
Connection mode Network
Conguration XML le
"GSDML-Vx.xx-Sartorius-PR5230-xxxxxx.xml"
Potential isolation yes
Cable type Twisted pairs, screened, e.g. patch cable CAT5 Autolink (stra-
ight or crossover)
Cable impedance
150 Ω
Cable length to HUB Max. 115 m
Certicate ProBus Nutzerorganisation e.V. for HMS Industrial Net-
works AB
Certicate No: Z10931
Note:
The IP address and the subnet mask are set at - [Fieldbus parameter] (see also
Chapters 7.15.5 and 12.2).
The XML le is stored on the CD (Fieldbus directory of the corresponding instrument)
supplied with the unit. The current le is also available for download via the Internet.
http://www.minebea-intec.com
Transmitter in eld housing PR 5230 4 Device installation
Minebea Intec EN-96