EasyManua.ls Logo

Mirion Technologies Lynx II DSA - General and Coincidence Acquisition Settings

Mirion Technologies Lynx II DSA
217 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 Web-Based Operations
92 Lynx II DSA User's Manual - 7096089
General
The General tab displays general acquisition information. The screen below shows the
general acquisition while in the PHA mode.
Figure 45: General Acquisition Settings
No. of Channels
The Input size determines the amount of acquisition memory in a number of channels that
the Lynx II assigns to the currently selected Input number. The value can range from 256
channels to 32768 channels. Each channel can store up to 2
32
-1 (approx. 4.29B) counts.
For Input #1, the Input Size represents the number of channels, or granularity, of the
digitized PHA value over the full energy range of the input signal.
For Input #2, the Input Size represents the number of channels, or sequential bins, into
which the time-correlated data (based on the current dwell time) will be stored.
Preset Options
Select from Real, Live, or None which disables the Preset Limits fields.
Preset Limits
Time preset can be Seconds, Minutes, Hours, or Days. Preset time can be in effect while
other preset conditions (i.e. Comp Preset) are active. Fractional values for the respective
unit range are accepted. Effective preset time can range from 0.1s to 42,948,000s.

Table of Contents

Related product manuals