EasyManuals Logo

Mirion Technologies Lynx II DSA User Manual

Mirion Technologies Lynx II DSA
217 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #29 background imageLoading...
Page #29 background image
Rear Panel Connectors
Lynx II DSA User's Manual 7096089 19
±200 V
±200 V dc output (range is –200 to +200 V)
All HVPS outputs are current limited and short-circuit protected.
CAUTION:
Incorrect connection and/or control settings of the HVPS for
your Detector can damage your Detector.
The High Voltage Outputs may be subject to the status and settings for a High Voltage
Inhibit signal for a preamplifier. These protective features might be defeated if the Inhibit
Signal input is improperly connected or configured.
ENERGY
Accepts energy signals from a detector/preamplifier; polarity is software selectable.
HV INH
High Voltage Inhibit signal input. Accepts the preamplifier's Inhibit signal to shut down
the HVPS if the detector is above its safe operating temperature; the polarity of this input
is software selectable.
TRP INH
Accepts inhibit signal from a transistor reset preamp; its function depends on the selected
mode, Automatic or Manual.
MCS
Accepts MCS input signals.
Advanced Panel
The Advanced panel provides the Input/Output connectors to be used for advanced
functions. For the connector's technical data, refer to the Specifications on page 135
Inputs
CHGR RDY
Accepts Sample Changer Ready signals; polarity is software selectable.
GATE for Coincidence /Anti-coincidence
Coincidence / Anticoincidence signal input; polarity is software selectable.
MCS SS
Accepts external signals to start/stop data acquisition if using MCS mode; two software
selectable modes (one for Start, Stop, or Start & Stop modes; the other for Suspend &
Resume mode).
CHAN ADV
MCS Channel Advance accepts external MCS channel advance signals.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mirion Technologies Lynx II DSA and is the answer not in the manual?

Mirion Technologies Lynx II DSA Specifications

General IconGeneral
BrandMirion Technologies
ModelLynx II DSA
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals