Chapter 5 Web-Based Operations
122 Lynx II DSA User's Manual - 7096089
Preset
The SCAs support both Live and Real time preset.
Preset Limit (mS)
The value must be entered in milliseconds. Range is 10 milliseconds to 2
32
-1 milliseconds.
Clock Source
The synchronization Clock Source is enabled only when External Sync Status is set to
Independent. See Hardware Option | External Synchronization.
If External Synchronization Status control is set to Independent, the Clock Source
setting then defines the clock source to be used when external synchronization is
Enabled. Options are Internal or External, as follows:
Internal means that the clock source is the internal 1 µS clock.
External means that the clock source is defined by the External Sync input at the
rear panel’s SYNC connector.
Mode
Defines how all of the SCAs are acquired and buffered - Automatic or Manual. Manual
indicates that you must invoke the latch and clear commands. Automatic indicates that the
software will invoke the latch and clear automatically; once latched the SCA counter
values are automatically buffered.
Automatic will cause the Lynx II to automatically buffer SCA data based on the
preset conditions. The buffer may be retrieved using the Lynx Programming
Libraries.
Manual will give control to the browser or third-party application. SCA data will
not be buffered, and the preset conditions are ignored because counting is based on
the control of the browser or third-party application.
Enable
Selecting the Enable option starts data collection.
Time Series
Select “On” to enable or “Off” to disable the plot. A view of the SCA counter data buffers
in an automatically updating time series plot.
Note: Enabling the Time Series will disable the DSO if it was enabled.
To expand the Time displayed, press the left-mouse button down and drag the cursor of
the region to expand. The plot will highlight the region. Release the left button.
To return to the normal display, double click the left-mouse button in the expanded region.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mirion Technologies Lynx II DSA and is the answer not in the manual?

Mirion Technologies Lynx II DSA Specifications

General IconGeneral
BrandMirion Technologies
ModelLynx II DSA
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals