Appendix A Specifications
136 Lynx II DSA User's Manual - 7096089
PHA SS
External PHA Start/Stop input connector, MCX. TTL compatible, software-selectable
logic polarity for active-high or active-low. Two basic modes are supported, pulse or level,
programmable via software.
Pulse mode – Requires TTL pulse, width > 50 ns, positive or negative edge-
triggered, software selectable. Requires that the acquisition has been armed through
the Start Acquire command.
Startthe first pulse received will begin acquisition storage. Subsequent pulses
are ignored.
Stopif the acquisition is already active, the first pulse received will stop
acquisition; subsequent pulses are ignored. No action is taken if the acquisition
was off when the first pulse was received.
Start and Stopthe first pulse received will begin acquisition storage. Second
pulse received will stop the acquisition. Subsequent pulses will be ignored.
Suspend/Resume modeRequires TTL level, active high or active low, software
selectable. Requires that the acquisition has been armed through Start command.
Active signal begins or resumes acquisition. Inactive signal suspends acquisition.
Either mode requires that the acquisition has been armed with start-acquire command.
MCS SS
External MCS Start/Stop input connector, MCX. TTL compatible, software-selectable
logic polarity for active-high or active-low. Two modes are supported, pulse or level,
programmable via software.
Pulse mode – Requires TTL pulse, width > 50 ns, positive or negative edge-
triggered, software selectable. Requires that the acquisition has been armed through
the Start Acquire command.
Startthe first pulse received will begin acquisition storage. Subsequent pulses are
ignored.
Stopif the acquisition is already active, the first pulse received will stop
acquisition; subsequent pulses are ignored. No action is taken if the acquisition was
off when the first pulse was received.
Start and Stop – first pulse received will begin acquisition storage. Second pulse
received will stop the acquisition. Subsequent pulses will be ignored.
Suspend/Resume modeRequires TTL level, active high or active low, software
selectable. Requires that the acquisition has been armed through Start command.
Active signal begins or resumes acquisition. Inactive signal suspends acquisition.
Either mode requires that the acquisition has been armed with start-acquire command.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mirion Technologies Lynx II DSA and is the answer not in the manual?

Mirion Technologies Lynx II DSA Specifications

General IconGeneral
BrandMirion Technologies
ModelLynx II DSA
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals