Appendix E Operation with Reset Preamps
178 Lynx II DSA User's Manual - 7096089
As shown below, the overload recovery time is approximately 32 µs, with the Lynx II Rise
Time set to 5.6 µs, Flat Top set to 0.8 µs, Coarse Gain of x11.32 , and using a Mirion
Model 2101 preamp and
60
Co source. For this example, the Inhibit Time should be set to
approximately 35 µs.
Figure 79: Digital Oscilloscope Waveform Showing Trapezoid Filter