Chapter 11 METER SETTING
11.4 Maintenance Setting
142 HORIBA
BACK LIGHT OFF TIME
The transfer time can be set. This is the time duration until the backlight is extinguished with
no key operation.
1. Press the ENTER key.
The setting is now available.
2. Press or key and set the time.
Setting range: 0 to 999 (min)
Setting 0 keeps the backlight in lit state.
3. Press the ENTER key.
The setting is made.
Pressing any key returns the display from power saving state to normal state.
CLEAR DATA MEMORY
This function deletes all the measurement results stored in the system memory.
1. Press the ENTER key.
The data memory is deleted.
The data stored in the compact flash is not deleted.
INITIALIZATION OF SETTINGS
The set values, calibration values, and the operator information can be cleared.
1. Press the ENTER key.
The system configuration is reset to the one of factory setting, and the power is turned
off.
2. Press ON or OFF button.
The power is turned on.
"16.6 Default settings" page 249