Chapter 5 Ion Measurement (F-53,55)
5.4 Changing Measurement SETUP Conditions
F-52, 53, 54, 55 83
● Setting the Ion type
The Ion type setting does not necessary for setting the Ion electrode model.
When the electrode model setting is not necessary or when the model is not freely input, the
Ion valency should be input by manual operation.
1. In the ION SETUP screen, select the ION SELECTION option.
2. Press ← or → key and select MANUAL.
Selecting MANUAL:
1. Press the ENTER key.
The display color of the input part of ION SELECTION changes, enabling the input.
2. Press ↑ or ↓ key and set the Ion valency (-2, -1, 1, 2).
3. Press F4 (RETURN).
The setting is reflected, and the display returns to the Measurement screen.
● Changing the measurement unit
1. In the ION SETUP screen, press ↑ or ↓ key and select the UNIT option.
Press ← or → key to select either g/L or mol/L.
2. Press F4 (RETURN) to reflect the setting. The display returns to the Measure-
ment screen.