System Connector Cable (W08) and Auxiliary Connector Cable
(W09) Replacement
Section Menu Section Contents Procedures Back Index
332
8
Removing the System Connector Cable (W08) or Auxiliary Power Cable (W09)
Refer to Figure 8.109 on p. 331.
To remove the system connector cable (3009652-01) or auxiliary power cable (V2 part 3303848-002), from the rear case:
1. Disassemble the case as described in Disassembling the Case (p. 181).
2. Remove the parameter bezel as described in Removing the Parameter Bezel (p. 311).
3. Remove the power PCB as described in Removing the Power PCB (A03) (p. 254).
4. If required, remove the system connector cable nut (806091-00), connector seal (805487-00), and O-ring seal (200060-011).
Remove the System connector cable.
5. If required, remove the auxiliary power cable nut (806091-00), connector seal (805487-00), and O-ring seal (200060-011).
Remove the connector.